 Preliminary Technical Data
FEATURES
Low Power, 5-Electrode Electrocardiogram (ECG) Analog Front End ADAS1000
APPLICATIONS
ECG: Monitor & Diagnostic Bedside Patient Monitoring, Portable Telemetry, Holter, AED, Cardiac Defibrillators, Ambulatory Monitors, Pace Maker Programmer, Patient Transport, Stress testing,
Biopotential signals in; digitized signals out 5 acquisition (ECG) channels + 1 driven lead Parallel ICs for up to 10+ Electrode measurements AC and DC Leads Off Detection Internal Pace Detection Algorithm on 3 leads Support for users own pace Thoracic Impedance Measurement (internal/external path) Selectable Reference Lead Scalable Noise vs. Power Control Low Power Operation from: 11mW (one lead), 14mW (3 leads), 19mW (all electrodes) Power Down Modes Lead or Electrode Data available Supports AAMI EC11:1991/(R)2001/(R)2007, EC13:2002/(R)2007, IEC60601-1 ed. 3.0b, 2005, IEC606012-25 ed. 1.0b, IEC60601-2-27 ed. 2.0, 2005 IEC60601-2-51 ed. 1.0, 2005 Fast Overload Recovery Low or High speed Data Output Rates Serial interface SPI(R)-/QSPITM-/DSP-compatible 56 lead LFCSP package (8mm x8mm) 64 lead LQFP package (10mm x10mm Body size)
GENERAL DESCRIPTION
The ADAS1000 measures electro cardiac (ECG) signals, thoracic impedance, pacing artifacts, lead on/off status and outputs this information in the form of a data frame supplying either Lead/Vector or Electrode data at programmable data rates. Its low power and small size make it suitable for portable, battery powered applications. The high performance also makes it suitable for higher end diagnostic machines. The ADAS1000 is designed to simplify the task of acquiring and ensuring quality ECG signals. Value-added cardiac postprocessing is executed externally on a DSP, microprocessor or FPGA. Several digital output options ensure flexibility when monitoring and analyzing signals. The ADAS1000 provides a low power, small data acquisition system for biopotential applications.
FUNCTIONAL BLOCK DIAGRAM
REFIN REFOUT Voltage Reference
CALIBRATION DAC
CAL_DAC _IO CM_IO
SHIELD DRIVE
RLD_SJ
RLD_FB RLD_OUT DRIVEN LEAD AMP
+ SHIELD
DRIVE AMP
VCM (1.5V)
-
+
RESPIRATION DAC
AC LEADS OFF DAC
COMMON MODE AMP
LEADS OFF DETECTION
MUXES
PACE DETECTION
CS SCLK STANDARD SERIAL INTERFACE
5 x ECG PATH
ELECTRODES x5
SHIELD DRIVE
AMP
ADC
FILTERS, CONTROL & INTERFACE LOGIC
SDI SDO DRDY GPIO0/MCS
EXT RESP 1
GPIO1/MSCLK
AMP
OPTIONAL
EXT RESP 2 EXT RESP 3
ADC
CLOCK GEN/OSC /EXTERNAL CLK SOURCE
XTAL IN XTAL O UT
GPIO2/MSDO GPIO3 CLK_IO
SECONDARY MASTER INTERFACE FOR PACE
RESPIRATION PATH
8.192MHz Cyrstal or Clock
Figure 1 Functional Block Diagram
Rev. Pr.D
Information furnished by Analog Devices is believed to be accurate and reliable. However, no responsibility is assumed by Analog Devices for its use, nor for any infringements of patents or other rights of third parties that may result from its use. Specifications subject to change without notice. No license is granted by implication or otherwise under any patent or patent rights of Analog Devices. Trademarks and registered trademarks are the property of their respective owners.
One Technology Way, P.O. Box 9106, Norwood, MA 02062-9106, U.S.A. www.analog.com Tel: 781.329.4700 Fax: 781.461.3113 (c)2011 Analog Devices, Inc. All rights reserved.
ADAS1000
GENERAL DESCRIPTION - CONTINUED
Preliminary Technical Data
Auxiliary features that aid in better quality ECG signal acquisition include: multi-channel averaged driven lead, selectable reference drive, fast over load recovery, flexible respiration circuitry returning magnitude and phase information, internal pace detection algorithm operating on 3 leads and option of AC or DC leads off detection. Since ECG systems span different applications, the ADAS1000 features a novel power/noise scaling architecture where the noise can be reduced at the expense of increasing power consumption. Signal acquisition channels may be shutdown to save power. Data rates can be reduced to save power. To ease manufacturing tests, development, as well as offer holistic power-up testing, the ADAS1000 offers a suite of dc and ac test excitation via the CAL DAC, CRC redundancy testing in addition to read-back of all relevant register address space. The ADAS1000 is available in two package options, either a 56 lead LFCSP or a 64 lead LQFP package and is specified over -40C to +85C temperature range.
Rev. Pr.D | Page 2 of 3
Preliminary Technical Data OUTLINE DIMENSIONS
8.10 8.00 SQ 7.90 0.60 MAX 0.60 MAX
43 42
ADAS1000
0.30 0.23 0.18
56 1
PIN 1 INDICATOR
PIN 1 INDICATOR
7.85 7.75 SQ 7.65
0.50 BSC
EXPOSED PAD
5.25 5.10 SQ 4.95
14 29
TOP VIEW 1.00 0.85 0.80 SEATING PLANE 12 MAX 0.80 MAX 0.65 TYP
0.50 0.40 0.30 0.05 MAX 0.02 NOM COPLANARITY 0.20 REF 0.08
28
15
BOTTOM VIEW 6.50 REF
0.25 MIN
FOR PROPER CONNECTION OF THE EXPOSED PAD, REFER TO THE PIN CONFIGURATION AND FUNCTION DESCRIPTIONS SECTION OF THIS DATA SHEET.
081809-B
COMPLIANT TO JEDEC STANDARDS MO-220-VLLD-2
Figure 2. 56 lead, 8mm x 8mm LFCSP
0.75 0.60 0.45 12.20 12.00 SQ 11.80
64 1 PIN 1 49 48
1.60 MAX
TOP VIEW
(PINS DOWN)
10.20 10.00 SQ 9.80
1.45 1.40 1.35
0.15 0.05
SEATING PLANE
0.20 0.09 7 3.5 0
16 17 32
33
0.08 COPLANARITY
VIEW A
VIEW A
ROTATED 90 CCW
0.50 BSC LEAD PITCH
0.27 0.22 0.17
051706-A
COMPLIANT TO JEDEC STANDARDS MS-026-BCD
Figure 3. 64 lead, 10mm x 10mm body size LQFP
ORDERING GUIDE
Model 1 ADAS1000BCPZ ADAS1000BSTZ EVAL-ADAS1000EBZU1
1
Temperature Range -40C to +85C -40C to +85C 1st Silicon Evaluation kit
Package Description 56 lead LFCSP 64 lead LQFP
Lead Finish TBD TBD
Package Option CP-56 ST-64
Z = Pb-free part.
(c) 2011 Analog Devices, Inc. All rights reserved. Trademarks and registered trademarks are the property of their respective owners. PR09660-0-1/11(PrD)
Rev. Pr.D | Page 3 of 3


▲Up To Search▲   

 
Price & Availability of ADAS1000

All Rights Reserved © IC-ON-LINE 2003 - 2022  

[Add Bookmark] [Contact Us] [Link exchange] [Privacy policy]
Mirror Sites :  [www.datasheet.hk]   [www.maxim4u.com]  [www.ic-on-line.cn] [www.ic-on-line.com] [www.ic-on-line.net] [www.alldatasheet.com.cn] [www.gdcy.com]  [www.gdcy.net]


 . . . . .
  We use cookies to deliver the best possible web experience and assist with our advertising efforts. By continuing to use this site, you consent to the use of cookies. For more information on cookies, please take a look at our Privacy Policy. X